Stainless steel isolation membrane dedicated pressure transmitter
Hotline:+8613330243969
outline structure
 
 

a.png

Technical Parameter
 
 

Pressure parameters

range


0MPa~0.1MPa…70MPa


overload

2 times the full-scale pressure

pressure type

Gauge pressure, absolute pressure

Electrical parameters

type of output

current mode

voltage source

power supply

9~30V DC

12~30V DC

output signal

4mA~20mA DC (two-wire system)

0/1V~5/10V DC (three wire system)

load resistance

≤ (U-9)/0.02 Ω (two-wire system)

≥ 10k (three wire system)

structure parameters

housing

stainless steel

sensor

316L stainless steel

sealing ring

VITON

protection grade

IP65

environmental condition

Media applicability

Various fluids that are non corrosive to 316L stainless steel and fluororubber

compensation temperature

0℃~70℃

operation temperature

-30℃~80℃

Storage temperature


-40℃~125℃


performance index

Accuracy

≤±0.5%FS

zero temperature

≤±0.02%FS/℃

Full temperature coefficient

≤±0.02%FS/℃

Long term stability

≤± 0.3% FS/year (maximum)

Electrical connection
 
 

cable color

two-wire

Three-wire

black

Power supply positive:+ V

Power supply positive:+ V

red

Signal:+ OUT

Output positive:+ OUT

white

empty

Public end: GND

The electrical connection method for the 4-20mA DC output of the two-wire system is shown in the following figure:
 
 

b.jpg

Electrical connection of transmitter with 3-wire 0/1-5VDC output
 
 




Hotline